The orthopedic market offers more than 200 hundred different femoral stems, some with scientific published results. The influence of stem geometry relatively to stress distribution on the cement mantle has been addressed by some authors. Stem geometry influences the fatigue mechanism process at the stem-cement interface and may cause aseptic loosening. This study presents a new cemented hip femoral concept. The study was based on finite element analysis and in vitro hip replacements. The numerical models allowed us to determine the cement mantle stresses of commercial stems that were compared with the ones of the new concept. The stem developed presents innovative collar and tip geometries and “organic” stem sections. The new cemented hip prosthesis reduces the average cement stresses around 25% relatively to the best commercial stems studied.
Design Of Medical Devices Conference Abstracts
A New Cemented Hip Prosthesis Concept
José A. Simões
José A. Simões
University of Aveiro
Search for other works by this author on:
António Ramos
University of Aveiro
Carlos Relvas
University of Aveiro
António Completo
University of Aveiro
José A. Simões
University of Aveiro
J. Med. Devices. Jun 2011, 5(2): 027503 (1 pages)
Published Online: June 3, 2011
Article history
Published:
June 1, 2011
Online:
June 3, 2011
Citation
Ramos, A., Relvas, C., Completo, A., and Simões, J. A. (June 3, 2011). "A New Cemented Hip Prosthesis Concept." ASME. J. Med. Devices. June 2011; 5(2): 027503. https://doi.org/10.1115/1.3587102
Download citation file: